Abstract

With current 30-MHz intravascular ultrasound systems, flowing blood may cause considerable backscatter which in real-time images is characterized by dynamic speckle. However, in a single intravascular ultrasound image (still-frame) the discrimination between arterial lumen and wall may be difficult due to the frozen intraluminal speckle, particularly in the presence of dissections. We compared subtraction, averaging and saline flush as methods to improve the discrimination between arterial lumen and wall in a single image. The real-time images served as gold standard. In 22 patients who underwent peripheral balloon angioplasty, ultrasound images obtained from 84 sites were examined. The sensitivity and specificity of detecting dissections were in the subtraction image 85% and 100%, in the averaged image 57% and 96%, and in the saline flush image 58% and 86%, respectively. Subtraction is a promising method to outline the irregular lumen in a single image.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.