Abstract

AbstractThe behavior of the Fe(II)/(III) redox system at a Au microelectrode ensemble (Au‐MEE) based on a solid composite by means of direct and cyclic voltammetric analysis (VA) is reported. With a simple electrode activation and sample preparation, the influence of dissolved organic substances was eliminated, providing highly sensitive results. The analytical signal was based on the maximum cathodic current (I) of the first derivative (dI/dE), and iron determination within the 0.002–0.04 mg L−1 range was studied. A sensitive LOD (3σ) value of 0.7 µg L−1 for total iron concentration was calculated; total iron determination in different waters was shown.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.