Abstract

A solventless analytical method is proposed for analyzing the compounds responsible for cork taint in cork stoppers. Direct sample introduction (DSI) is evaluated as a sample introduction system for the gas chromatography-mass spectrometry (GC–MS) determination of four haloanisoles (HAs) in cork samples. Several parameters affecting the DSI step, including desorption temperature and time, gas flow rate and other focusing parameters, were optimized using univariate and multivariate approaches. The proposed method shows high sensitivity and minimises sample handling, with detection limits of 1.6–2.6ngg−1, depending on the compound. The suitability of the optimized procedure as a screening method was evaluated by obtaining decision limits (CCα) and detection capabilities (CCβ) for each analyte, which were found to be in 6.9–11.8 and 8.7–14.8ngg−1, respectively, depending on the compound. Twenty-four cork samples were analysed, and 2,4,6-trichloroanisole was found in four of them at levels between 12.6 and 53ngg−1.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.