Abstract
A simple and compact instrument for NO 2 measurement by laser-induced fluorescence (LIF) technique with a pulsed solid state laser and a multi-pass excitation system was developed and optimized for several conditions. As a result of laboratory experiment, the limit of detection (LOD) reached 94 pptv for 60 s integration. It was thought that an LIF instrument with this LOD value would be capable of quantifying sub-ppbv NO 2 in unpolluted marine atmosphere. As the second step, a field test of the instrument was conducted in the marine atmosphere at Cape Hedo, Okinawa Island, Japan, in summer 1999. Intercomparison between the LIF instrument and a chemiluminescence detector with a photolytic converter (PLC-CL) was also made in this test. Consequently, the LIF instrument was shown to be of practical use for measuring NO 2 in clean maritime air.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.