Abstract
A rapid and sensitive method based on direct infusion-nano-electrospray ionization mass spectrometry (DI-nESI-MS) has been developed for the detection and quantification of ciprofloxacin and its metabolites in human saliva. Saliva samples were collected after the oral administration of 500 mg ciprofloxacin tablets. Internal standard (IS), tamoxifen, was added to the collected samples, and then diluted with the ionization solvent, centrifuged and filtered. An aliquot of 4 μL of the filtrate was loaded into a nanospray (NS) capillary. The NS capillary was then fitted into an off-line ion source and the instrument was operated to acquire a two-minute run by applying a voltage of 1000 V (positive-ion detection mode). Quantification of ciprofloxacin relied on the ratio of its peak intensity to the IS peak intensity. The DI-nESI-MS method was validated and provided satisfactory precision with relative standard deviation ranging from 0.39 to 7.48 % and accuracy with relative error ranging from -2.12 to 9.72 %. The calibration curve showed good linearity (r2) > 0.999 over the concentration range of 10−4000 ng/mL. These results verify the effectiveness of the DI-nESI-MS method for monitoring of ciprofloxacin and its metabolites in human saliva samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.