Abstract

In this flow-injection system for direct determination of lithium in serum by atomic absorption spectroscopy, the 10-microL sample is manually injected into a continuously flowing non-segmented stream of de-ionized water, which is pumped, via a dispersion tube, to the spectrometer's nebulizer. Controlled dispersion of the sample zone, before it is introduced into the nebulizer, produces the required sample dilution. Effects of varying the length of the dispersion tube, the flow rate, and the sample size were studied. Analytical readout is obtained, in the form of transient peaks, 5 s after sample injection. It is necessary to include physiological concentrations of sodium and potassium in the standard because each of these cations enhances the lithium absorbance signal. Analytical recovery (98.5 to 101%) and CV (about 2%) are good, and results compare well with those obtained by aspiration of prediluted samples (n = 121, r = 0.99).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.