Abstract

A method has been developed for the direct determination of lithium in serum and whole blood by graphite furnace atomic absorption spectrometry using tantalum carbide coated graphite tubes in combination with ammonium nitrate matrix modification. The maximum permissible charring temperature for lithium was raised from 1050 to 1500 °C and the lifetime of the tube was also prolonged for ca. 200 more firings using the tantalum carbide coated graphite tube compared with the conventional pyrolytic graphite coated graphite tube. Matrix interferences from chlorides present in human blood samples were eliminated using the proposed method, which has been used for the determination of lithium in serum and whole blood of normal healthy individuals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.