Abstract

Herein, direct determination of small RNAs is described using a functional-polymer modified genosensor. The analytical strategy adopted involves deposition by electropolymerization of biotinylated polythiophene films on the surface of miniaturized, disposable, gold screen-printed electrodes, followed by the layer-by-layer deposition of streptavidin, and then biotynilated capture probes. A small RNA (miR-221) target was determined via the impedimetric measurement of the hybridization event in a label-free and PCR-free approach. Under optimized conditions, the limit of detection (LOD) was 0.7 pM miR-221 (15% RSD). The genosensor was applied for determination of miR-221 in total RNA extracted from human lung and breast cancer cell lines, discriminating between the cancer-positive and -negative cells, without any amplification step, in less than 2h.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.