Abstract

A novel method was developed for the direct determination of trace quantities of rare earth elements (REEs) in high purity erbium oxide dissolved in nitric acid by inductively coupled plasma mass spectrometry (ICP-MS) in this work. The mass spectra overlap interferences arose from Er matrix on the neighbouring and monoisotopic analytes of 165Ho(100) and 169Tm(100) were eliminated by adjusting instrumental peak resolution value from 0.7 to 0.3 amu. The matrix suppression effect of Er on the ion peak signals of REEs impurities was effectively compensated with spiking In as internal standard element. The limit of quantitation (LOQ) of REEs impurities was from 0.0090 to 0.025 μg g −1, the recoveries of spiked sample for REEs were found to be in the range of 90.3–107% through using the proposed method and relative standard deviation (R.S.D.) varied between 2.5% and 6.7%. The novel methodology had been found to be suitable for the direct determination of trace REEs impurities in 99.999–99.9999% high purity Er 2O 3 and the results obtained from this method keep in good agreement with that acquired from high resolution ICP-MS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.