Abstract

We propose a measurement method for sensitive and label-free detections of virus-like particles (VLPs) using color images of nanoplasmonic sensing chips. The nanoplasmonic chip consists of 5×5 gold nanoslit arrays and the gold surface is modified with specific antibodies for spike protein. The resonant wavelength of the 430-nm-period gold nanoslit arrays underwater environment is about 570 nm which falls between the green and red bands of the color CCD. The captured VLPs by the specific antibodies shift the plasmonic resonance of the gold nanoslits. It results in an increased brightness of green pixels and decreased brightness of red pixels. The image contrast signals of (green - red) / (red + green) show good linearity with the surface particle density. The experimental tests show the image contrast method can detect 100-nm polystyrene particles with a surface density smaller than 2 particles/µm2. We demonstrate the application for direct detection of SARS-CoV-2 VLPs using a simple scanner platform. A detection limit smaller than 1 pg/mL with a detection time less than 30 minutes can be achieved.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.