Abstract
Fused-silica columns of 0.22 mm I.D. packed with normal small-particle high-performance liquid chromatography packing material were studied. The shape of the column end allows direct connection to a combined electron impact—chemical ionization (EI—CI) ion source. To inject small volumes a syringe-loaded micro-sample injector is used with the time splitting technique. Mass spectra obtained from easily vaporized compounds are similar to, but display less thermal decomposition than, normal EI spectra. Mass spectra of underivatized native mono- and disaccharides give useful structural information. Results from the analysis of cardiac glucosides are also shown. The system provides high-efficiency chromatography with a universal detector, suitable for the analysis of non-volatile and labile compounds.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.