Abstract
Microcystins are a group of structurally similar cyclic heptapeptide hepatotoxins and tumor promoters, produced by cyanobacteria. A microbore liquid chromatography electrospray ionization ion-trap mass spectrometry (LC-ESI-ITMS) method has been developed which is capable of separating and detecting trace amounts of microcystin variants in environmental samples. Extracted water sample was loaded onto a LC trapping column and, using a column switching technique, the compounds of interest were back-flushed onto a 1-mm LC column. Structural elucidation was achieved using ion-trap with tandem mass spectrometry in the data dependent scan mode. Collision-induced dissociation to MS 3 allowed tentative identification of these cyclic peptides. Full-scan LC-ESI-MS mass spectrum was obtained when 250 pg of the authentic compound was injected onto the HPLC column, which represents the detection limit for microcystin-LR. This study demonstrated that LC-ESI-ITMS is a reliable and sensitive technique for analysing trace levels of microcystins.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.