Abstract

A digital phase-sensitive detector was designed and built to circumvent the relatively slow, ∼1-ms response time of commercial lock-in amplifiers. A Z80A single-board computer with a suitably programmed EPROM and analog-to-digital and digital-to-analog converters is used. The response time, ∼40 μs, is primarily limited by the throughput of the analog-to-digital converter.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.