Abstract
Ziock et al.'s [1] recent Monte Carlo study of a proposed ``full-volume'' Compton Imaging Camera concluded that simultaneously locating a Compton scatter event's multiple interaction points within a single large scintillator crystal might be possible at 1 mm spatial resolution using a coded aperture mask sandwiched between two light guides and coupled to a position sensitive photomultiplier (PSPMT) to record the output light pattern. The method promises high efficiency at a relatively low cost. They are currently developing a lower resolution prototype using a large cubic scintillator (25.4 cm/side) whose masked face will be tiled with 25 Hamamatsu H9500 PSPMTs (6,400 outputs). XIA has contracted to develop and produce the readout electronics, which present several significant design challenges, including capturing all 6,400 anode outputs individually, with single photon sensitivity, in a compact format that will fit behind the tiled PSPMTs. 10,000 event/sec operation is desired, as is a cost of less than about $50/channel.In our approach, each PSPMT front end integrates the 256 anode signals and 8-1 multiplexes them to 32 differential outputs that are digitized in a PXI card using 4 octal 50 MHz ADCs. The multiplexers run at 8 MHz, sampling each anode at 1 MHz, which becomes the image frame rate. The ADC signals are demultiplexed and digitally filtered to extract the number of photons in each pixel in the full 2-D image. The design has been completed and built and is undergoing evaluation tests at the single PSPMT level.