Abstract

To acquire isotropic high-resolution, signal-to-noise ratio (SNR) efficient, 3D encoded diffusion-weighted MRI data. Multiple Fourier encoded slabs were combined into one full volume, using a generalized autocalibrating partially parallel acquisitions (GRAPPA) accelerated diffusion-weighted 3D multislab echo planar imaging (EPI) sequence with 2D phase navigation and in-plane motion correction. Reconstructed data with 1.5-mm(3) nominal resolution is presented and shown under the influence of motion and with variable slab thicknesses. The SNR efficiency between diffusion-weighted 3D multislab EPI and DW 2D ss-EPI is compared. Finally, a 1.3-mm(3) full brain scan with 45 diffusion directions is presented. Diffusion-weighted 3D multislab EPI has been presented as an alternative sequence for high-resolution and high-SNR full brain coverage diffusion studies. Compared with the gold standard 2D diffusion-weighted single-shot Echo Planar Imaging, the SNR efficiency is significantly higher.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.