Abstract
Abstract A diffusion scrubber-ion chromatographic (DS-IC) instrument has been characterized and employed for the measurement of trace levels of gaseous HCl in the atmosphere. The instrument operates with a temporal resolution of 5 min and the detection limit is estimated to be 5 pptv. Collection efficiencies for HCl with two identical diffusion scrubbers were 28±2% and 20±2%, respectively, at a sampling flow rate of 2 SLPM. Instrument response decreases with increased relative humidity. An equation, correction factor=2.45 × 10−7 × %r.h3 + 1.00, is used to correct for the relative humidity dependency. The instrument was tested in ambient air studies by measuring background mixing ratios between 0.02 and 0.5 ppbv at a suburban sampling site. Calibration of the instrument was carried out with a novel source of gaseous HCl based on sublimation of ammonium chloride.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.