Abstract
High performance π-shaped capacitive micro-machined ultrasonic transducers (CMUTs) are presented with improved linearity and sensitivity. The π-shaped CMUT is composed of an anchor membrane, a circular piston electrode plate, and a bridging layer. The proposed differential π-shaped CMUT can provide differential output signals and further increase the sensitivity by about 6 dB. Both π-shaped CMUTs were designed and verified in a TSMC 1P6M 180nm CMOS MEMS technology. Experimental results show that the differential π-shaped CMUT not only provides superior linearity under 0–0.6 MPa pressure, but also exhibits 1.6 and 2.46 times in sensitivity compared to the conventional CMUT architecture.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.