Abstract

The 3ω technique has been applied to measure thermal properties of thin films and suspended wires. In this article, we report the implementation and technical design of a 3ω system to measure the thermal conductivity in small dimension structures. Alternating current is applied to a line heater that dissipates power and varies its resistance. The 3ω voltage drop across the heater provides information that determines the thermal properties of the heater or of the adjacent material. A differential bridge is used to subtract the first harmonic voltage and systematic noise. We introduce the use of a synchronous detector associated with the differential bridge to obtain the in-phase and out-of-phase components of the third harmonic signal. The combination improves the bridge method for a broad range of frequency analysis that permits either thermal conductivity and/or heat capacity measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.