Abstract

A diaphragm-type pressure transducer with a sputtered platinum film strain gauge (sensing film) has been designed and fabricated. The various steps followed to prepare thin film strain gauges on the diaphragm are described. M-bond 450 adhesive (Measurements Group, USA) has been employed as the insulating layer. A detailed procedure to cure this layer is given. A d.c. sputtering method is employed to prepare the platinum films. This paper also includes details of the strain gauge pattern and its location on the diaphragm. A description of the output characteristics and overall behaviour of the platinum thin film pressure transducer is reported.