Abstract

The establishment of typical diagnostic reference levels (DRLs) values according to the complexity indices (CIs) for hepatic chemoembolisation (HC), iliac stent placement (ISP) and femoropopliteal revascularisation (FR) is reported in this study. To estimate patients' stochastic effects, effective dose was calculated through dose area product (DAP) values of this study and E/DAP conversion factors derived from the literature. Data for DAP, Reference Air Kerma (Ka,r) and fluoroscopy time (FT) were collected for 218 patients and CIs were assigned to each procedure to extract DRLs. To estimate effective dose, conversion factors and DAP values were used for seven IR procedures. DRL values for DAP were 141, 130 and 28Gy*cm2 for HC, ISP, and FR, respectively. The corresponding DRL values for Ka,r were 634.6, 300.1 and 112.0mGy, and for FT were 15.3, 12.4 and 17.9min, respectively. CIs in interventional radiology are a useful tool for the optimisation of DRLs since they contribute to patient's doses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.