BackgroundAccurate diagnosis of anemia by community workers using a point-of-care device is a challenge. The objective of the study was to establish the diagnostic accuracy of point-of-care devices for detecting anemia in community settings.MethodsIt was diagnostic accuracy study with cross-sectional design on adult patients attending the outpatient department of rural/ urban health centres of Medical colleges from India. The index tests were HemoCue, TrueHb, Massimo’s device and spectroscopic device, compared against autoanalyzer (gold standard). Accuracy was expressed by sensitivity, specificity, likelihood ratios, predictive values, area under the curve (AUC) and levels of agreement. For the diagnostic accuracy component, 1407 participants were recruited with a minimum of 600 for each device. An additional 200 participants were considered to elucidate the performance of devices in different weather conditions.ResultsHemoCue and TrueHb performed better than Massimo and spectroscopic devices. Detection of anemia by technicians was similar between TrueHb and HemoCue (AUC 0.92 v/s 0.90, p > 0.05). Community workers performed better with Hemocue for detecting anemia compared to TrueHb (AUC 0.92 v/s 0.90, p < 0.05). For detection of severe anemia, accuracy of TrueHb was significantly better with technicians (AUC 0.91 v/s 0.70; p < 0.05) and community workers (AUC 0.91 v/s 0.73; p < 0.05).HemoCue showed a bias or mean difference (95%CI) of 0.47 g/dl (0.42, 0.52) for all values, and 0.92 g/dl (0.82, 1.03) for severe anemia. For TrueHb, it was − 0.28 g/dl (− 0.37, − 0.20) for all readings, and 0.06 g/dl (− 0.52, 0.63) for severe anemia. TrueHb appeared to be more consistent across different weather conditions, although it overestimated Hb in extreme cold weather conditions.ConclusionFor detection of anemia, True Hb and HemoCue were comparable. For severe anemia, True Hb seemed to be a better and feasible point-of-care device for detecting anemia in the community settings.

Highlights

  • Accurate diagnosis of anemia by community workers using a point-of-care device is a challenge

  • Some devices that have the potential for use in field settings in India include invasive devices and non- invasive devices

  • The second phase, a cross sectional study assessed the accuracy of devices in different weather conditions

Read more

Summary

Introduction

Accurate diagnosis of anemia by community workers using a point-of-care device is a challenge. The objective of the study was to establish the diagnostic accuracy of point-of-care devices for detecting anemia in community settings. Hemoglobin (Hb) estimation is the cornerstone for diagnosis of anemia. The choice of methods depends on site of use (clinical/ community settings), availability of resources, validity of the device, the user (technicians/ health workers/ nurses), and population (adults/ blood donors/ children/ pregnant women), to name a few. Some devices that have the potential for use in field settings in India include invasive devices (require finger pricking for taking blood - HemoCue, TrueHb) and non- invasive devices (do not require pricking -Massimo’s device and AJO spectroscopic device). The findings are not very consistent [12, 15,16,17]

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.