Abstract

Assessment of urinary pterins is proposed as a rapid method for recognition of the variants of hyperphenylalaninemia. This is achieved by means of oxidation of pterins by iodine in acidic and alkaline solutions and then by high performance liquid chromatography on a cation-exchange column with fluorimetric detection. In biopterin-synthetase deficiency, only neopterin accumulated; in dihydropteridine-reductase (DHPR) deficiency and in phenylketonuria, high levels of pterins are found, but BH 4 levels, absent in the former and high in the latter, allow a differential diagnosis. Phenylalanine loads in the controls also lead to increased elimination of pterins, but with a pattern different from that found in phenylketonuria. This method can be used before dietary treatment and thus can be proposed for all newly detected hyperphenylalaninemic babies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.