With the rapid development of laser medicine, there are higher requirements placed on the performance of optical components in various medical systems. This paper is aimed at exploring the critical optical devices of medical equipment for treating periodontitis and gingivitis. The cathode sputtering method was used to produce the wide-angle short-wave pass filter, and a hundreds grid fastness test was conducted to detect the occurrence of film peeling. Considering the results of SEM, transmission spectrum, and stress test of the sample, an analysis was conducted as to the cause of poor bonding force for the film. By increasing the amount of argon gas and adjusting the baking temperature, the problem of film peeling was resolved. Besides, a short-wave pass filter film with good bonding and low roughness was obtained to meet the requirements of laser medical equipment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.