Abstract

BackgroundThe smartphone-based whole slide imaging (WSI) system represents a low-cost and effective alternative to automatic scanners for telepathology. In a previous study, the development of one such solution, named scalable whole slide imaging (sWSI), was presented and analyzed. A clinical evaluation of its iOS version with 100 frozen section samples verified the diagnosis-readiness of the produced virtual slides.ObjectiveThe first aim of this study was to delve into the quantifying issues encountered in the development of an Android version. It should also provide insights into future high-resolution real-time feedback medical imaging apps on Android and invoke the awareness of smartphone manufacturers for collaboration. The second aim of this study was to further verify the clinical value of sWSI with cytology samples. This type is different from the frozen section samples in that they require finer detail on the cellular level.MethodsDuring sWSI development on Android, it was discovered that many models do not support uncompressed camera pixel data with sufficient resolution and full field of view. The proportion of models supporting the optimal format was estimated in a test on 200 mainstream Android models. Other factors, including slower processing speed and camera preview freezing, also led to inferior performance of sWSI on Android compared with the iOS version. The processing speed was mostly determined by the central processing unit frequency in theory, and the relationship was investigated in the 200-model simulation experiment with physical devices. The camera preview freezing was caused by the lag between triggering photo capture and resuming preview. In the clinical evaluation, 100 ThinPrep cytology test samples covering 6 diseases were scanned with sWSI and compared against the ground truth of optical microscopy.ResultsAmong the tested Android models, only 3.0% (6/200) provided an optimal data format, meeting all criteria of quality and efficiency. The image-processing speed demonstrated a positive relationship with the central processing unit frequency but to a smaller degree than expected and was highly model-dependent. The virtual slides produced by sWSI on Android and iOS of ThinPrep cytology test samples achieved similar high quality. Using optical microscopy as the ground truth, pathologists made a correct diagnosis on 87.5% (175/200) of the cases with sWSI virtual slides. Depending on the sWSI version and the pathologist in charge, the kappa value varied between .70 and .82. All participating pathologists considered the quality of the sWSI virtual slides in the experiment to be adequate for routine usage.ConclusionsLimited by hardware and operating system support, the performance of sWSI on mainstream Android smartphones did not fully match the iOS version. However, in practice, this difference was not significant, and both were adequate for digitizing most of the sample types for telepathology consultation.

Highlights

  • With the data quality and speed improvements of automated microscopes and whole slide scanners [1,2], telepathology has become a major component in pathology labs [3,4]

  • Using optical microscopy as the ground truth, pathologists made a correct diagnosis on 87.5% (175/200) of the cases with scalable whole slide imaging (sWSI) virtual slides

  • All participating pathologists considered the quality of the sWSI virtual slides in the experiment to be adequate for routine usage

Read more

Summary

Introduction

With the data quality and speed improvements of automated microscopes and whole slide scanners [1,2], telepathology has become a major component in pathology labs [3,4]. The high financial cost of whole slide imaging (WSI) solutions, and especially the up-front portion, has not seen a significant reduction after years of maturity, limiting its penetration into developing countries and regions or remote hospitals in the developed world [10,11,12]. As supplements for these situations, where limited manual operating is preferred over expensive automation, manual and low-cost alternatives of automated WSI have been studied and developed [13,14]. A clinical evaluation of its iOS version with 100 frozen section samples verified the diagnosis-readiness of the produced virtual slides

Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.