Abstract
A novel sample preparation method, vial wall sorptive extraction (VWSE), which uses a vial whose internal wall is coated with polydimethylsiloxane (PDMS), was developed. The method was applied to the determination of progesterone in human serum sample. Human serum sample (0.5 mL) spiked with progesterone- 13C 2 was pipetted into the VWSE device and vortex mixing was performed for 30 min. Then, the serum sample was removed and the vial rinsed with purified water. Fifty microliter of methanol as liquid desorption (LD) solvent was pipetted into the VWSE device and vortex mixing was performed for 10 min. Then, the extract was analyzed by liquid chromatography-tandem mass spectrometry (LC-MS/MS). The correlation coefficient ( r) of the calibration curve over the concentration range of 0.5–200 ng mL −1 was 0.999. The limit of detection (LOD) and the limit of quantification (LOQ) were 0.1 and 0.5 ng mL −1, respectively. The relative recoveries were 97.9% (RSD: 4.4%, n = 6) and 102.8% (RSD: 1.1%, n = 6) for progesterone spiked at 5 and 50 ng mL −1, respectively. This simple, accurate, sensitive, and selective analytical method is applicable to the trace analysis of a minute amount of sample.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.