Abstract
Total decay energy spectrometry with cryogenic detectors is a promising technique for radionuclide analysis of α-emitting nuclides. The radioactive sample is embedded in the detector absorber, and the total decay energy for each disintegration is measured as a temperature elevation. We are developing this technique with metallic magnetic calorimeters (MMCs). The main condition of this technique is a detection efficiency close to unity. However, some α-emitting nuclides emit intense γ-rays that can partially escape from the absorber. So a feasibility study for several nuclides has been carried out based on Monte Carlo simulations of the detection efficiency and numerical calculations of the expected energy resolution to identify the radionuclides that can potentially be measured. Furthermore, an MMC prototype has been built and tested. The total decay energy spectrum of Po-210 was measured, and a FWHM energy resolution of 1.25 keV at 5.4 MeV was obtained and a Gaussian width at half maximum of 0.827 (5) keV. The baseline FWHM energy resolution is 207 eV, consistent with the resolution obtained on low-energy L X-rays in the same spectrum at 14 keV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.