Abstract
We developed a position-sensitive gamma-ray detection system consisting of a double-sided silicon strip detector (DSSD) and a 25-segmented germanium detector (25-SEGD). Two major applications of the system are gamma-ray imaging (Compton camera) and linear polarization measurement for gamma rays emitted from oriented nuclei. Customized electronics were developed in order to handle multi-channel signals of both the DSSD and the 25-SEGD. Images for a 133 Ba-based compound source in a square shape with areal dimensions of 1.5 × 1.5 mm 2 are presented. Comparison between experimental images and a Monte Carlo simulation yielded the overall imaging resolution within 1 cm for the present system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.