Abstract

A calibration facility based on the gravimetric method has been developed at NRLM for small mass flow rates of gases. This calibration facility establishes a primary gas flow measurement standard in Japan with an expanded maximum uncertainty of ±0.12 percent for small mass flow rates from 0.1 g / min to 50 g / min for nitrogen gas. Five sonic Venturi nozzles were tested using this facility and the results showed that they can work well as a flow transfer standard for small mass flow rates. These sonic Venturi nozzles can be used to establish flow transfer standards within an expanded uncertainty of ±0.15 percent.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.