Abstract

We present new diagnostics for use in optical laser pump - X-ray Free Electron Laser (XFEL) probe experiments to monitor dimensions, intensity profile and focusability of the XFEL beam and to control initial quality and homogeneity of targets to be driven by optical laser pulse. By developing X-ray imaging, based on the use of an LiF crystal detector, we were able to measure the distribution of energy inside a hard X-ray beam with unprecedented high spatial resolution (∼1 μm) and across a field of view larger than some millimetres. This diagnostic can be used in situ, provides a very high dynamic range, has an extremely limited cost, and is relatively easy to be implemented in pump-probe experiments. The proposed methods were successfully applied in pump-probe experiments at the SPring-8 Angstrom Compact free electron LAser (SACLA) XFEL facility and its potential was demonstrated for current and future High Energy Density Science experiments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.