Abstract

Ultra-violet (UV) light source is one of the important components of ultraviolet photoelectron spectroscopy (UPS). A Helium direct current discharge ultraviolet light source is mainly composed of anode, quartz capillary, gas inlet pipe, differential exhaust assembly and heat sink. At 80Pa working pressure and 30mA discharge current, the test sample shows 1nA photoelectron current. Photoelectron current increases proportional with discharge current. The results show that the He DCDUV light source is working well and can be used as the light source in UPS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.