Abstract
Development of a dissociation-enhanced lanthanide fluoroimmunoassay (DELFIA) for measuring leptin, a satiety hormone of appetite control, was conducted in sandwich assay format exploiting a microplate immobilized with an anti-leptin antibody and another antibody raised against leptin and tagged with an europium chelate. In the leptin DELFIA of this study, amounts of antibody coated to the microplate and of the bioconjugate for the second immune reaction were optimized as 0.5μg and 200ng per well, respectively. When plotted in double-logarithmic scale, a linear relationship of y (log10 response signal)=0.6023× (log10 leptin concentration)+3.4084 (r(2)=0.9646) was obtained at the leptin concentrations of 0.01─50ng/mL with the limit of detection of 0.01ng/mL. Individual leptin concentrations in various samples were well convergent to the calibration curve of the current assay. When applied to the measurement of leptin in a rat serum, the present assay was found quite effective and was competitive to a commercial sandwich-type ELISA.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have