Abstract

We present a spectrometer for sensitive absorption measurements in liquids across broad spectral bandwidths. The spectrometer combines the unique spectral properties of incoherent supercontinuum light sources with the advantages of cavity ring-down spectroscopy, which is a self-calibrating technique. A custom-built avalanche photodiode array is used for detection, permitting the simultaneous measurement of ring-down times for up to 64 different spectral components at nanosecond temporal resolution. The minimum detectable absorption coefficient was measured to be 3.2 × 10(-6) cm(-1) Hz(-1/2) at 527 nm. We show that the spectrometer is capable of recording spectral differences in trace levels of blood before and after hemolysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.