Abstract
For on-site analysis of surface materials on the Moon, planets, and small bodies and for the monitoring of air quality in crewed spacecraft, we have developed a portable gas chromatograph (GC) equipped with a ball surface acoustic wave (SAW) sensor. In this study, we fabricated a 10 cm cube GC that implements the forward flush method using two metal micro-electro-mechanical-system columns coated with different stationary phases in microchannels fabricated by wet etching and diffusion bonding of stainless-steel plates. Using this GC, we succeeded in analyzing 10 kinds of gas within 10 min. In addition, for the application of the ball SAW GC on the ground, we also developed a palm-sized GC with a single metal capillary column and used it in the analysis of the headspace gas of sake. We showed that the ratio of peak areas differed among odorants depending on the brand and brewing process of sake.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.