Abstract

This imaging system aims at recording images of the core size and shape of an imploding deuterium-tritium (DT) microballoon on LMJ inertial confinement fusion (ICF) experiments. Image acquisition is difficult due to the harsh surrounding created by the fusion reaction, which affects system specifications. This one is made of a scintillator, an optical relay, and a CCD camera shielded from the surrounding. The system was tested on different facilities at CEA/DIF, where a spatial resolution of 120 μm was achieved and gamma dose up to 20 rad effects were measured. Setup and performed test are described.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.