Abstract

Polycyclic aromatic hydrocarbons (PAHs) are widespread environmental contaminants formed during incomplete combustion or pyrolysis of organic material. The reliable quantification of PAH in airborne samples is still difficult, costly, and time-consuming due to the use of offline techniques, including long sampling on filters/adsorbents, laboratory extraction, purification, and concentration steps before analysis. To tackle these drawbacks, this work focused on the development of a fully automatic gas chromatograph (GC) equipped with a flame ionization detector (FID) and a sample preconcentration unit (PC) for gas sampling. This instrument was validated under laboratory-controlled conditions in the range 0–10 ng for 18 PAH. The chromatographic separation was rather satisfactory except for two PAH pairs, which were quantified together. For all compounds, the peak areas increased perfectly with the gaseous PAH concentration (R2 > 0.98), without any significant memory effect between two consecutive analyses. Considering a gaseous sample volume of 1 L, the extrapolated limits of detections (LOD) were in the range 19.9–62.6 ng/m3, depending on the PAH. Its analytical performances were then compared to those of the offline reference UHPLC-fluorescence method, widely used for airborne PAH monitoring. This was also compared with the very few portable or continuously operating instruments.