Abstract

To develop an improved dosemeter to assess photon and beta exposures of the eye lens, and in response to issues surrounding the preferred values of Hlens to be used for guiding operational radiation protection, a programme of re-optimization of the current PHE thermoluminescence dosemeter has been performed. In particular, refinements of the filter located in front of the sensitive 7LiF:Mg,Cu,P element have been considered, so that the dose response characteristics of the device provide a better and more conservative estimate of risk. The investigation was performed using the Monte Carlo modelling software MCNP5, to produce a final design that featured a filter containing a 9.5 mm diameter polypropylene hemisphere truncated to a maximum thickness of 3.0 mm. The responses of this design in photon and electron fields are presented here, contrasted against those of the existing PHE eye dosemeter, with respect to the operational quantity Hp(3,E,θ) and both current and suggested values for the absorbed dose per fluence risk profile for the lens of the eye.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.