Abstract

High sensitivity and high resolution is desired in such technologies as neutron radiography. However, the contamination of gamma photons in neutron images decreases the accuracy of neutron radiography. To solve this problem, we developed an event-by-event based neutron imaging system that can selectively detect neutrons. The developed neutron imaging system consists of an Li–ZnS(Ag) scintillator plate optically coupled to a flat panel photomultiplier tube (FP-PMT) with a light guide. Scintillation light emitted from the Li–ZnS(Ag) by the interaction with neutron-induced particles is used to calculate the position based on the center of mass calculations. The spatial resolution of the neutron imaging detector is ∼2.3 mm FWHM, and the sensitivity for 252Cf at 2 cm from the source with 2-cm-thick polystyrene is 20 cps/MBq. Background count fraction from 60Co gamma photons was 0.2 %. For various types of neutron absorption phantoms, high-contrast and high-resolution neutron images are obtained with the developed neutron imaging detector using a252Cf neutron source.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call