Abstract

We have developed a procedure for the fabrication of Er–Ni liquid alloy ion sources. The source is tested and analyzed in a MicroBeam 150 focused ion beam system. Our experimental results show that an Er2+ target current of 95–100 pA was produced, representing 50% of the total target current. The ion emission current–voltage slope is ∼36 μA/kV. Source lifetimes are generally larger than 20 h. The alloy oxidizes quickly once it is exposed to air. This must be minimized to ensure proper source performance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.