Abstract

A compact electron cyclotron resonance ion source has been developed for heavy-ion medical facilities. The beam intensity and stability were considerably improved by recent modifications on three points (length of sextupole, cooling system for the extraction electrode, and position of the Einzel lens). Initial results of C4+ beam tests show that an intensity of 180 eμA can be routinely obtained with simple tuning. The best record was 220 eμA for C4+, which meets the medical requirements. Throughout these tests, CH4 gas was used with 0.1 cc/min and the extraction voltage was fixed at 25 kV. Results on beam emittance and long-term stability are also briefly discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.