In this study, a manganese ferrite (MnFe2O4) based dispersive solid phase extraction method (MF-DSPE) was proposed for the determination of lead (Pb) ions in hibiscus tea by double slotted quartz tube flame atomic absorption spectrometry (SQT-FAAS). The MnFe2O4 nanoparticles used as sorbent were produced in high yields using a green and friendly microwave synthesis procedure. The limit of detection (LOD) was found to be 14.2 µg/L under the optimal conditions. Recovery experiments were performed to assess the applicability of the developed method using hibiscus tea as test sample matrix. High recovery results were obtained by applying the matrix matching calibration method. Thanks to the developed MF-DSPE-SQT-FAAS method, the detection sensitivity for lead was improved by about 108-folds according to the comparison of the slopes of the calibration plots of the FAAS and MF-DSPE-SQT-FAAS systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.