Abstract

The proposed method demonstrates that the osmium(VIII) forms complex with 2-NBATCH from 0.8molL−1 HCl at room temperature. The complex formed was extracted in 10mL of chloroform with a 5min equilibration time. The absorbance of the red colored complex was measured at 440nm against the reagent blank. The Beer's law was obeyed in the range of 5–25μgmL−1, the optimum concentration range was 10–20μgmL−1 of osmium(VIII) as evaluated by Ringbom's plot. Molar absorptivity and Sandell's sensitivity of osmium(VIII)-2NBATCH complex in chloroform is 8.94×103Lmol−1cm−1 and 0.021μgcm−2, respectively. The composition of osmium(VIII)-2NBATCH complex was 1:2 investigated from Job's method of continuous variation, Mole ratio method and slope ratio method. The interference of diverse ions was studied and masking agents were used wherever necessary. The present method was successfully applied for determination of osmium(VIII) from binary, ternary and synthetic mixtures corresponding to alloys and real samples. The validity of the method was confirmed by finding the relative standard deviation for five determinations which was 0.29%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.