Abstract

Pathogenic bacteria can cause the outbreaks of disease and threaten human health, which stimulates the development of advanced detection techniques. Herein, a specific and sensitive electrochemical biosensor for Gram-negative bacteria was established based on the conductive polymer with artificial muscle properties. Theeffective recognition was achieved through the specific carbohydrate-carbohydrate interaction between gluconamide and lipopolysaccharide. Theapplication of impulse voltage enhances the efficiency of recognition and shortens the detection time through the temporary deformation of the electrode surface, with a limit of detection (LOD) of1 × 100CFU/mL anda linear range of 1 × 100 - 1 × 106CFU/mL for Escherichia coli (E. coli). In addition tothe merits of low cost, high efficiency, and rapidity, the developedlabel-free electrochemical biosensor can also be applicable for other Gram-negative bacteria, owning promising potential in the application of portable devices and paving a potential way for the construction of electrochemical biosensors.