Abstract

The 4DM-PET experiment aims to develop an innovative detector module for PET applications. The device is composed of a continuous LYSO scintillator crystal coupled to two SiPM matrices on its top and bottom surfaces. The peculiarity in using a continuous crystal is the ability to reconstruct the Depth of Interaction of the 511keV photons generated by the annihilation of the positrons emitted by the radiotracer. A first small prototype module has been built. It is composed of a LYSO crystal, 20×20×10mm3 coupled on larger surfaces with arrays of SiPMs (4×4pixel, 4×4mm2 each). The lateral faces of the crystal slab are black painted to avoid reflection of light. The detector is read by custom designed Front-End ASICs. We have scanned the detector in the three coordinates with pencil beams produced by collimated radioactive sources in order to investigate the module spatial resolution capabilities. The results achieved in terms of Depth of Interaction (DOI) reconstruction and Point Spread Function (PSF) at different positions on the detector surface are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.