Abstract
Abstract A solid-state pH sensor was fabricated using a transparent conductive tin oxide film on a glass substrate. The coating of the glass substrate was achieved by a novel simple chemical vapor deposition (CVD) procedure. The response time of the pH sensor was substantially reduced when a thin graphite film was deposited onto the tin oxide conductive film. The sensor slope was found to increase as the temperature of the solution was increased. The performance of the sensor was investigated in the pH range from 0.3 to 11.0. A straight-line calibration graph was achieved throughout the whole range tested, especially when the solution temperature was 80 ℃. The working pH range was found to decrease on the expense of the lower range as the temperature was decreased. Results obtained by the suggested sensor compares very well with conventional pH electrodes where the square of the correlation coefficient was 0.999.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.