Abstract

A novel detector for flow injection analysis has been developed based on the optical beam deflection induced by the chemical reaction heat. A new flow cell was designed, in which the lower part is filled with CCU and an HCl solution is flowed above the CCU phase. An NaOH solution is injected into the flow line to react with the HCl. The neutralization heat is transferred to the CCU phase, thus generating a temperature gradient. This temperature gradient induces a refractive index gradient which, in turn, generates a deflection of a probe beam that is passed through the CCl4 phase. The effects of the flow rate and the length of the cell on the deflection signals have been investigated. The deflection signal was found to be proportional to the concentration of NaOH in the range of 0.05 to 0.5 mol/l by using the flow cell. The detection limit of the NaOH concentration was about 0.05 mol/l. Another type of the flow cell is also discussed. c]Changchun, p]China. c]Minami-Ohsawa, Hachioji, Tokyo o]192-03, p]Japan

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.