Abstract

Non-destructive measurements of low-intensity charged particle beams are particularly challenging for beam diagnostics. At the Heavy Ion Accelerator Facility in Lanzhou (HIRFL), beams with weak currents below 1 µA are often provided for experiments. The detection of such low beam current is below the threshold of typical standard beam current transformers. Therefore, a low-intensity monitoring system is developed by using a sensitive capacitive pick-up (PU) and low-noise electronics. This device measures beam currents by digitally analyzing the amplitude of the PU signals using a homodyne detection scheme. During lab tests, the amplitude nonlinearity is <0.5% in the operational range of 1 nA-45 µA and the amplitude resolution is 0.94 nA. At present, four measurement systems for low beam currents are installed at HIRFL for the monitoring of standard operating conditions with low beam currents below 1 µA. After an absolute calibration with a Faraday cup, it can be used for accurate beam intensity measurement with a current resolution of about 1 nA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.