Abstract

A high-sensitivity multidimensional gamma-spectrometer is being developed within the shallow underground laboratory at Pacific Northwest National Laboratory (PNNL, USA). The system consists of two broad energy germanium detectors, inside a low-background shield, fitted with a cosmic veto system. The detector has advanced functionality, including operation in single or combined detector mode, with reductions in the cosmic background of 49.6% and Compton suppression of 6.5%. For selected radionuclides this provides increased peak identification, reductions in uncertainty of 27.6% and MDA improvements of 52.7%. The design uses commercially off-the-shelf components to provide a powerful solution for low-level nuclear measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.