Abstract

A light-weight portable neutron survey meter was developed using a mixed organic gas counter for dose management at nuclear power plants and accelerator facilities. This survey meter, NSN31041, is ~2 kg in weight and W160×H250×L300 mm(3) in size, which is capable of measuring neutron ambient dose equivalent rate from thermal to 15 MeV neutrons. The neutron energy response of the survey meter is evaluated using continuous energy neutron sources of (252)Cf, (241)Am-Be, thermal neutrons generated from a graphite pile loading a (252)Cf source, concrete-moderated neutrons of (241)Am-Be source and D(2)O-moderated neutrons of (252)Cf source. The measured response data show very good agreement with neutron ambient dose equivalent within a 50 % deviation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.