Abstract

We have developed a high-efficiency high-resolution particle-induced x-ray emission (PIXE) system employing a von Hamos-type crystal spectrometer for a chemical state identification of trace elements in environmental samples. The energy resolution of the system was determined to be about 0.05% through the observation of Si Kalpha(1,2) x rays (1.74 keV) from elemental silicon. The throughput efficiency of the system was also evaluated quasitheoretically to be 1.6x10(-7) counts/incident proton for Si Kalpha(1,2) emission. To demonstrate a chemical state analysis using the high-resolution PIXE system, Si Kalpha(1,2) and Kbeta x-ray spectra for SiC, Si(3)N(4), and SiO(2) were measured and compared. The observed chemical shifts of the Si Kalpha(1,2) peaks for SiC, Si(3)N(4), and SiO(2) relative to elemental silicon were 0.20, 0.40, and 0.55 eV, respectively. The tendency of these shifts were well explained by the effective charges of the silicon atoms calculated by a molecular orbital method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.