Abstract

An analytical method based on a dual capillary gas chromatographic technique combining the advantages of GasPro PLOT and a non polar narrow bore WCOT column was developed for the analysis of air samples containing C2–C9 NMHCs. A refocusing step was not required due to the fast heating rate of the sample preconcentration trap and the resolving power of the PLOT column for C2 and C3 NMHCs. Water had to be removed from the air samples to avoid plugging of the columns if the initial GC oven temperature was below ambient temperature. To dry air samples, a scrubber and a cryogenic technique were employed. The interferences caused by carbon dioxide were reduced by purging the loaded sample preconcentration trap with helium. The dual column system was compared to a method employing a refocusing device and a single narrow bore WCOT column. Both systems provided a high degree of precision. However, the dual column approach was superior to the single column system due to better resolution of low molecular weight components.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.